System Power Requirements

Calculating Your Power Needs
To determine whether an auxiliary power supply card is required for your system, refer to the Options Power-Requirement Worktable at right. Then, list the quantity of the particular DBK option in use, and multiply the quantity by the mW power consumption indicated.

Place the resulting totals in the last column, add them, and list the overall total at the bottom of the column. If the overall total is equal to or less than the appropriate number indicated in the Available Power Chart below, the application does not require an auxiliary power supply. If your overall total exceeds the amount indicated in the chart below, then one or more DBK32As are required.

DBK Options Power-Requirement Worktable

DBK
Options
Voltage References Quantity mW Total
+15 VDC -15 VDC +5 VDC
DBK1 0 0 0 x 0 =
DBK2† 18 mA 18 mA 5 mA x 565 =
DBK4 95 mA 80 mA 25 mA x 2750 =
DBK5† 2 mA 2 mA 15 mA x 135 =
DBK7 14 mA 8 mA 18 mA x 420 =
DBK8 15 mA 15 mA <1 mA x 455 =
DBK9 21 mA 16 mA <1 mA x 560 =
DBK11A 0 0 0 x 0 =
DBK15 16 mA 16 mA <1 mA x 485 =
DBK16 37 mA 32 mA <1 mA x 1040 =
DBK17 30 mA 30 mA <1 mA x 905 =
DBK18 36 mA 36 mA <1 mA x 1085 =
DBK20 0 0 <10 mA x 50 =
DBK21 0 0 <10 mA x 50 =
DBK23** 0 0 <2 mA x 10 =
DBK24** 0 0 <2 mA x 10 =
DBK25 0 0 <2 mA x 10 =
DBK41 0 0 0 x 0 =
DBK42** <1 mA <1 mA <1 mA x 35 =
DBK43A** <1 mA <1 mA <1 mA x 35 =
DBK44 <1 mA <1 mA 60 mA†† x 330 =
DBK45 52 mA 52 mA <1 mA x 1565 =
DBK46 20 mA 20 mA 400 mA x 2600 =
DBK48† 25 mA 25 mA <1 mA x 750 =
DBK50** <1 mA <1 mA <1 mA x 35 =
DBK51** <1 mA <1 mA <1 mA x 35 =
DBK55† 30 mA 20 mA 40 mA x 950 =
DBK65† <25 mA <25 mA <1 mA x 755 =
DBK70** <1 mA <1mA <1 mA x 35 =
DBK80 25 mA 25 mA <1 mA x 755 =
DBK81 35 mA 35 mA <2 mA x 1060 =
DBK82 60 mA 60 mA <2 mA x 1810 =
DBK83 60 mA 60 mA <2 mA x 1810 =
DBK84 60 mA 60 mA <2 mA x 1810 =
DBK85† 25 mA 25 mA <1 mA x 755 =
DBK90 40 mA 40 mA 40 mA x 1400 =
Column Total

** Module with internal power supply, powered separately
† Not used with LogBook
†† 60 mA value is based on 30 mA for each of two 5B modules. This value will be higher if using 5B module 5B38 (200 mA for each 5B38), or if using 5B39 (170 mA for each 5B39).

A/D Available Power Chart

Available PowerProduct

Available mW

LogBook/300, /360* 5000
DaqBook/2000A, /2000E, /2000X 15000
DaqBook/2001, /2005 10000
DaqOEM/2001, /2005 10000
DaqBook/2020 7500
DaqBoard/2000 Series 2250
DaqLab/2000 Series 10000
DaqScan/2000 Series 10000

* LogBook does not function with DBK2 and DBK5

Additional Information

IOtech e-NewsCall MeDBK OptionsDBK Selection TableModule-to-Module ConnectionSoftwareTech Manuals

Get Acrobat Reader!

[ HOME | PRODUCTS | TECH SUPPORT | CONTACT | SEARCH/MAP | ABOUT IOTECH | SHOP ONLINE ]

IOtech PC-Based Data Acquisition (DAQ) and Instrumentation - PDF Format
® Copyright 2008, IOtech
Privacy PolicyTerms and Conditions of Sale

IOtech PC-based products (USB, PCI, Ethernet, GPIB) and signal conditioning options capture waveforms and measure most physical parameters including temperature, vibration, strain, velocity, acceleration, position, as well as common voltage, current, power, and data logging. IOtech's solutions are used in environments such as in-vehicle, OEM, embedded, industrial, aircraft, aerospace, laboratory, refineries, power generation, medical, and semiconductor facilities.